KEY POINTS
- The study retrospectively replanned 20 consecutive patients treated with non-hippocampal-sparing whole-brain RT using 30 Gy in 10 fractions and dual-arc VMAT. A C# script using the Monaco v6.2.2 API automated image import, template application, iterative optimization, dose normalization, DVH evaluation, and plan scoring.
- All 20 automated plans completed without script failure or manual mid-process intervention. Plans were normalized so that 95% of the whole-brain PTV received the full prescription, and all met institutional clinical dose criteria.
- Target coverage was essentially equivalent. Mean PTV D98 was 29.38 Gy manually versus 29.27 Gy automatically (p=0.092), while D2, D0.03cc, and V105% also showed no statistically significant differences.
- OAR performance was mixed. Automated plans reduced mean bilateral parotid dose from 6.75 to 5.67 Gy (p<0.001) and cochlear mean dose from 28.20 to 26.87 Gy (p=0.019), with parotid improvement seen in 19/20 patients and cochlear improvement in 15/20.
- Manual planning performed better for the optic nerves: D0.03cc was 28.74 Gy manually versus 30.15 Gy automatically (p=0.002). Most other conventional OAR dose metrics did not differ significantly.
- The automated plans produced higher modeled NTCP for several structures, including hippocampi, brainstem, eyes, and optic nerves. However, the absolute modeled probabilities were generally very low, including <0.2% for several organs, and the authors caution against overinterpreting relative percentage changes in model-derived risks.
- Recorded planning time decreased from 805.7 to 56.4 minutes, but this comparison is heavily biased because manual timestamps included queue and idle time rather than active dosimetrist work. Delivery time, monitor units, segment number, and total plan-quality score were essentially unchanged; the overall score was 47.66 versus 47.79 of 54.25 points.
CLINICAL TAKEAWAY
A scripted Monaco workflow can reliably generate a clinically acceptable baseline WBRT plan without manual intervention, potentially standardizing planning and enabling batch processing. It should currently be viewed as a dosimetrist-assist tool rather than a replacement for expert review, particularly because optic nerve sparing remained inferior and workflow-time savings were not measured as true hands-on labor.