KEY POINTS
- This multi-institutional study evaluated electron FLASH dosimetry systems across six participating institutions in North America and Europe, including University Hospital Zurich, Aarhus University Hospital, the University of Maryland, the University of Nebraska, Ohio State University, and Varian Medical Systems. Investigators characterized ionization chambers, diamond detectors, plastic scintillators, beam current transformers, radiochromic film, alanine, and luminescent dosimeters. Measurements were conducted using modified clinical linacs producing predominantly 16-MeV ultra-high-dose-rate electron beams, with additional 6-MeV measurements at one institution.
- The investigated platforms generated dose-per-pulse values exceeding those routinely encountered in conventional radiotherapy. Modified C-series systems achieved approximately 1.2 Gy/pulse at the isocenter, with average dose rates reaching approximately 215 Gy/s at 180 Hz. Modified TrueBeam systems produced >1.0 Gy/pulse and >200 Gy/s at the isocenter. Typical pulse duration was approximately 4.2 μs. These conditions create substantial challenges for detectors affected by charge recombination, saturation, and limited temporal response.
- Conventional ionization chambers demonstrated substantial dose-per-pulse-dependent underresponse. At the University of Nebraska, an Advanced Markus chamber remained approximately linear up to 0.4 Gy/pulse, but collection efficiency decreased to approximately 84% at 1.2 Gy/pulse. Measurements performed at Varian demonstrated saturation beginning at 0.2 Gy/pulse, with efficiency falling to approximately 36% at 7 Gy/pulse. These findings show that conventional chamber readings may substantially underestimate absorbed dose when used outside their validated ultra-high-dose-rate operating range.
- Dedicated FLASH ionization chambers extended the measurable range but remained dependent on beam energy. The PTW FLASH chamber demonstrated linearity up to approximately 0.9 Gy/pulse at 6 MeV, whereas saturation began around 0.4 Gy/pulse at 16 MeV, with approximately 91% collection efficiency at 2.3 Gy/pulse. These differences indicate that detector performance cannot be generalized across electron energies or assumed to remain linear throughout the full FLASH-relevant dose-per-pulse range.
- Plastic scintillators and specialized diamond detectors showed favorable responses under the tested conditions. The HYPERSCINT RP-100 scintillator demonstrated 97.7% efficiency at 1.4 Gy/pulse, while the Exradin W2 scintillator achieved 98.7% efficiency at 3.6 Gy/pulse. The PTW flashDiamond detector exhibited near-linear responses up to approximately 1.0 Gy/pulse for both 6- and 16-MeV beams, while the microDiamond detector remained linear to approximately 1.9 Gy/pulse before saturation. These systems also offer the potential for real-time or pulse-resolved measurements.
- Passive dosimetry provided an independent reference but lacked immediate feedback. Radiochromic film, alanine pellets, thermoluminescent dosimeters, and optically stimulated luminescent dosimeters generally demonstrated limited dose-rate dependence within their evaluated ranges. Alanine maintained approximately 96.7% efficiency at 1.4 Gy/pulse. Across the investigated pulse repetition frequencies, measured beam output varied by no more than 1.6%, while detector orientation typically changed response by approximately 1%. However, passive dosimetry requires calibration, delayed readout, and detector-specific correction procedures.
- No single detector fulfilled all measurement and quality-assurance requirements. The authors propose using radiochromic film for spatial reference dosimetry, alanine for independent absorbed-dose verification, scintillators or FLASH-capable diamond detectors for time-resolved measurements, and beam current transformers for nonperturbing beam monitoring. Because detector systems were tested using different institutional beams, energies, geometries, and calibration procedures, the results do not constitute a standardized head-to-head performance ranking. No patient-level efficacy or toxicity outcomes were investigated.
CLINICAL TAKEAWAY
Reliable electron FLASH dosimetry requires a combination of independent reference measurements and validated real-time monitoring, because conventional detectors can substantially underrespond at high dose per pulse. Specialized scintillators, diamond detectors, FLASH ionization chambers, and beam current transformers offer complementary capabilities, but each must be calibrated within its intended operating range. These experimental findings support detector-specific commissioning and quality assurance rather than a universal FLASH dosimetry solution.