Machine learning predicts prostate VMAT gamma passing rates with less than 1% average error

In 1,247 prostate VMAT plans, machine-learning models predicted gamma passing rates with approximately 0.9-percentage-point average error, without external validation.

KEY POINTS

  • This retrospective single-institution study analyzed 1,247 prostate VMAT treatment plans generated between January 2022 and December 2024 using Pinnacle version 16.2 and three beam-matched Elekta Versa HD linear accelerators. Pretreatment patient-specific QA was performed using the Delta4 diode-array system, with locally normalized 3%/3 mm gamma passing rate (GPR) as the prediction target. The institution considered GPR ≥98% clinically acceptable, making prediction accuracy near this threshold particularly relevant.
  • Nine machine-learning regression algorithms were evaluated using anatomical and plan-complexity features. Predictors included total monitor units, modulation complexity score, aperture variability, MLC leaf travel, small-aperture metrics, and the percentage overlap between PTV and pelvic organs at risk. The dataset was divided into 997 training plans (80%) and 250 holdout plans (20%), using stratified sampling to preserve the GPR distribution. Model development incorporated ten-fold cross-validation and hyperparameter optimization within the training cohort.
  • Tree-based ensemble models achieved the lowest prediction errors. In the reported holdout evaluation, Extra Trees achieved a mean absolute error (MAE) of 0.8906 percentage points, compared with 0.9054 for Random Forest and 0.8974 for Gradient Boosting. Random Forest achieved the lowest reported mean squared error in that evaluation (0.0157), followed by Extra Trees (0.0159). Simpler methods, including linear regression, decision trees, and neural networks, generally showed slightly higher errors.
  • Most holdout predictions were within three percentage points of the measured GPR. In the detailed uncertainty analysis, 98.4% of Extra Trees predictions and 98.0% of Random Forest predictions fell within ±3 percentage points of measured values. Extra Trees achieved a reported MAE of 0.8663 (95% CI 0.7520–0.9982) in this analysis, compared with 0.8925 (95% CI 0.7724–1.0328) for Random Forest. Pairwise comparisons between the three leading ensemble models were not statistically significant after Holm correction (adjusted p=1.0), so numerical rankings should not be interpreted as proven algorithmic superiority.
  • Plan modulation and anatomical complexity were important predictors of measured QA performance. Frequently influential features included aperture-area variability, edge metrics, modulation complexity, monitor units, and PTV overlap with the bladder or rectum. These relationships are physically plausible because smaller apertures, greater modulation, and more demanding target–OAR geometry increase delivery complexity. However, feature importance does not establish causality, and correlated complexity metrics can affect variable rankings.
  • Low average prediction error does not establish reliable detection of clinically unacceptable plans. The measured GPR distribution was strongly concentrated near high passing rates, with most plans exceeding the institutional 98% threshold. As a result, models could achieve favorable overall error statistics while having limited evidence for identifying rare low-GPR failures. The study primarily evaluated continuous GPR prediction rather than sensitivity, specificity, or negative predictive value for detecting clinically important QA failures.
  • External validation and prospective clinical governance remain essential. All plans originated from a single planning system, accelerator platform, detector configuration, and locally normalized gamma-analysis protocol. No external or temporal validation cohort was available, and the models did not incorporate delivery logs, EPID images, or independent dose recalculation. The investigators recommend prospective evaluation before using prediction results to modify measurement-based patient-specific QA requirements.

CLINICAL TAKEAWAY

Machine-learning models predicted measured gamma passing rates for prostate VMAT plans with average absolute errors near one percentage point in a large single-institution dataset. These results suggest potential for early identification of complex plans and prioritization of QA resources, but they do not establish that the models can reliably detect rare clinically unacceptable plans. External validation, failure-focused performance assessment, and prospective safety evaluation are required before reducing conventional measurement-based QA.

SOURCE

Medical Dosimetry

Browse more research Suggest a correction