Stereoscopic X-rays achieve submillimetric setup residuals during spine and pelvic bone SBRT

In 92 bone metastases, iterative stereoscopic X-ray guidance reduced setup residuals to 0.3 mm, with strong agreement against CBCT for spine and pelvic targets.

KEY POINTS

  • This retrospective single-center study from UZ Brussel, Belgium, evaluated a combined image-guidance and motion-monitoring workflow in 71 consecutive patients with 92 bone metastases treated with SBRT between April 2025 and July 2026. Targets included 57 spinal lesions (62%), 21 pelvic lesions (23%), six shoulder/proximal humerus lesions, and eight distal rib or sternal lesions. Most targets received five-fraction SBRT, commonly using 6–8.5 Gy per fraction, with CTV-to-PTV margins of 3–5 mm and a 1.5-mm PRV expansion around nearby critical structures.
  • The workflow combined stereoscopic X-ray registration, CBCT verification, and continuous optical/thermal surface guidance. Patients underwent initial surface-guided positioning, followed by iterative stereoscopic X-rays using ExacTrac Dynamic to correct six-degree-of-freedom positioning errors. CBCT provided volumetric verification, while additional stereoscopic X-rays were acquired at predefined gantry positions during treatment. Surface monitoring operated continuously between imaging acquisitions, with correction thresholds generally set at 1.5 mm/1.5° for X-ray monitoring and 2 mm/2° for surface guidance.
  • Stereoscopic X-ray and CBCT registrations demonstrated statistical equivalence for spinal and pelvic targets. Using two one-sided tests with strict equivalence limits of ±0.5 mm and ±0.5°, agreement was demonstrated across cervical, thoracic, lumbar, and pelvic anatomical regions (p<0.01). Successful initial agreement within 1 mm and 1° was achieved in all cervical and thoracic spine cases and all but one lumbar case, with 95% success for pelvic targets. Equivalent performance was not established for several shoulder and rib measurements, emphasizing the importance of anatomical location.
  • Iterative X-ray positioning substantially reduced residual setup errors. Across 250 fractions with three setup X-ray acquisitions, mean three-dimensional residual displacement decreased from 1.8 ± 2.1 mm to 0.3 ± 0.2 mm before CBCT. A typical three-image positioning sequence required 64 ± 35 seconds, including registration review and couch corrections, while the institutional CBCT workflow required approximately 173 ± 63 seconds for acquisition, registration, and correction. Individual stereoscopic acquisitions took approximately 2 seconds, compared with 60–85 seconds for CBCT acquisition and reconstruction.
  • Most intrafraction positioning checks remained within strict tolerances, but clinically relevant outliers occurred. Approximately 95% of stereoscopic X-ray verifications for spinal and pelvic targets remained within 1.5 mm and 1.5° across all monitored directions. Nevertheless, out-of-tolerance measurements were identified in 69 verifications across 40 fractions and 23 patients, with maximum deviations reaching 2.5 mm laterally and 3.4° in pitch for lumbosacral targets. Cervical spine cases showed rotational deviations up to 2.0° in roll, despite thermoplastic immobilization.
  • Surface guidance correlated strongly with internal bony motion only in selected directions. Compared with stereoscopic X-rays, Pearson correlation coefficients were r=0.872 laterally, r=0.881 for roll, and r=0.763 for yaw, but decreased to r=0.536 longitudinally and r=0.312 vertically. Respiratory motion and surface deformation contributed to poorer agreement and unnecessary alerts, particularly for thoracic and lumbar targets. Sustained surface-guidance deviations exceeding 2 mm/2° occurred in approximately 16% of spinal fractions, requiring additional image verification.
  • The results support anatomical site-specific monitoring rather than a universal image-guidance strategy. Stereoscopic registrations were less reliable for ribs and shoulders because of motion, overlapping anatomy, or limited bony landmarks, while peripheral long-bone and periarticular targets were excluded. CBCT was used as the reference standard despite not representing an absolute positional ground truth. The study did not assess delivered-dose accumulation, clinical local control, radiation myelopathy, or whether smaller PTV margins could be adopted safely.

CLINICAL TAKEAWAY

Combining stereoscopic X-rays with CBCT and surface guidance enabled highly reproducible positioning and frequent intrafraction verification during spine and pelvic bone SBRT, with mean residual setup errors reduced to approximately 0.3 mm. Surface monitoring provided useful additional information but was less reliable for respiratory-motion-related displacement and could not replace internal imaging. The findings support tailored image-guidance workflows and frequent position verification, but do not independently validate further reduction of PTV or spinal cord PRV margins.