120° short-arc CBCT preserves breast registration while cutting imaging dose by one-third

Reducing breast CBCT acquisition from 200° to 120° maintained submillimeter average registration differences while reducing central imaging dose by 33%.

KEY POINTS

  • The study evaluated short-arc CBCT for tangential breast IGRT using both an anthropomorphic phantom and clinical images from 30 consecutive whole-breast patients. The standard 200° CBCT acquisition was compared with reconstructed short-arc acquisitions down to 120°.
  • Phantom testing showed that central angle mattered as much as acquisition length. With a 120° arc, central angles around 130–140° best preserved the breast, chest-wall, and body-contour information needed for automatic registration.
  • In the phantom, reducing the acquisition angle to 120° maintained registration differences within approximately 0.15 mm for 3-axis and 0.55 mm for 6-axis matching. Performance deteriorated sharply only with very short acquisitions around 100°.
  • In clinical images reconstructed to 120°, mean registration difference versus standard CBCT was 0.64 mm for 3-axis matching, with 84.7% within 1 mm. For 6-axis registration, the mean was 0.85 mm, with 71.3% within 1 mm.
  • At 120°, all 3-axis registrations were within 2 mm, while 96% of 6-axis registrations were within 2 mm. These discrepancies remained smaller than the several-millimeter setup margins commonly used for whole-breast treatment.
  • Reducing the arc from 200° to 120° lowered measured central imaging dose by 33.4%. Acquisition time also decreased from approximately 33 seconds to 20 seconds, potentially making single-breath-hold CBCT more practical during DIBH.
  • Clinical short-arc datasets were predominantly reconstructed from full-acquisition projection data rather than prospectively acquired in patients, although phantom testing found close agreement between reconstructed and genuinely acquired short-arc images. The findings also cannot be assumed to apply to soft-tissue-dominant sites such as prostate or head and neck.

CLINICAL TAKEAWAY

For tangential whole-breast IGRT, a carefully selected 120° CBCT arc may meaningfully reduce imaging dose and scan time while preserving the anatomy required for positioning. Implementation should remain site-specific because acceptable acquisition angles depend on matching method, anatomy, and institutional setup margins.