KEY POINTS
- This retrospective single-institution study from South Korea investigated whether treatment-day cone-beam CT (CBCT) images could be predicted using planning CT and a fraction-specific cumulative planned-dose map. The dataset included 29 patients with lung cancer, 46 planning sequences, and 658 CBCT scans, including 17 patients who received an additional boost plan. Evaluation used 612 pairs of consecutive CBCT acquisitions and patient-level three-fold cross-validation, ensuring that images from the same patient were not divided between training and testing.
- Four image-generation architectures were compared using the same planning CT and dose-conditioning inputs: a denoising diffusion probabilistic model (DDPM), U-Net regression, ResViT, and pix2pixHD. The dose input represented the cumulative planned dose scaled to the treatment position, rather than actual deformably accumulated delivered dose. Evaluation had two components: static image reconstruction using structural similarity (SSIM) and mean absolute error (MAE), and longitudinal anatomical-change preservation using ChangeDice and gradient MAE.
- U-Net achieved the best conventional image-reconstruction performance. Whole-image SSIM was 0.71 ± 0.04 with U-Net versus 0.66 ± 0.05 with DDPM, while corresponding MAE values were 84.1 ± 20.0 versus 97.8 ± 21.8 HU. ResViT and pix2pixHD achieved intermediate static performance, with SSIM values of 0.70 and 0.68, respectively. These results indicate that deterministic regression generated images more closely resembling individual reference CBCT scans according to conventional image-quality metrics.
- The model ranking reversed when longitudinal anatomical change was evaluated. In the high-dose region, DDPM achieved ChangeDice 0.67 ± 0.06, compared with 0.34 ± 0.17 for U-Net, 0.33 ± 0.18 for ResViT, and 0.47 ± 0.15 for pix2pixHD. DDPM also achieved lower high-dose-region gradient error than U-Net (52.9 ± 8.4 versus 61.0 ± 12.1 HU). Thus, the model with better static similarity performed substantially worse when predicting where anatomical changes occurred between fractions.
- Multimodal language-model-derived conditioning produced only a small improvement within the diffusion framework. The investigators used Gemma 3 27B to generate fixed global and spatial representations from planning CT and dose information, which were then supplied to the diffusion model. Combining both representations increased high-dose-region ChangeDice by 0.011 (95% CI 0.002–0.020) compared with baseline DDPM. The improvement was consistent across cross-validation folds and two random seeds, but the absolute effect was small.
- Control experiments did not establish that the improvement was specifically attributable to the language model or patient-specific information. A mismatched-prior model performed worse than the correctly conditioned model during training comparisons. However, when the trained combined model received another patient's representation at inference, change overlap increased further by 0.019 (95% CI 0.014–0.024) while static image fidelity deteriorated. No control supplied equivalent numerical treatment information without the language model, preventing attribution of the observed benefit to language-model reasoning or patient-specific anatomical understanding.
- The framework remains a technical proof of concept rather than a validated adaptive radiotherapy tool. Each cross-validation model was trained on only 15–16 patients, all data originated from one institution and treatment platform, and independently generated diffusion images introduced sampling variability into the longitudinal comparisons. No direct validation of predicted tumor boundaries, OAR geometry, accumulated dose, or replanning decisions was performed. The authors explicitly distinguish the proposed prediction framework from actual treatment-day imaging and do not propose replacing CBCT acquisition.
CLINICAL TAKEAWAY
Predicting longitudinal anatomical evolution during lung radiotherapy requires different evaluation metrics from those used to judge the appearance of a single synthetic CBCT image. Diffusion models preserved inter-fraction change substantially better than deterministic regression, despite lower static image similarity, while language-model conditioning provided only a small and non-specific improvement. Clinical application will require anatomical and dosimetric validation before predicted changes can guide adaptive replanning decisions.