Halcyon shortens treatment delivery for selected sites compared with TrueBeam

In 60 paired plans and 120 timing assessments, Halcyon reduced delivery time for breast, head and neck, and brain/head treatments.

KEY POINTS

  • This two-phase study from Central West Cancer Centre, Australia, compared Halcyon and TrueBeam across six treatment-site categories. Phase 1 comprised 60 patients, with ten paired plans per category: head and neck, chest/abdomen, prostate with or without nodal irradiation, brain/head, free-breathing breast, and palliative treatment. Phase 2 evaluated treatment delivery times in 120 patients, with 20 per category. All plans used Eclipse version 16.1, with 6-MV FFF beams for Halcyon and 6- or 10-MV beams for TrueBeam.
  • Both platforms achieved clinically acceptable target coverage, with a small statistical advantage for Halcyon. Mean PTV coverage was 97.88 ± 1.82% with Halcyon versus 97.63 ± 1.98% with TrueBeam (p=0.037), an absolute difference of 0.25 percentage points. All plans met the departmental minimum target-coverage requirements. Although statistically significant, the small difference does not establish a clinically meaningful improvement in tumor control.
  • Composite OAR dose metrics favored Halcyon, particularly for selected anatomical sites. Across the paired plans, pooled mean OAR dose was 14.31 Gy with Halcyon versus 14.68 Gy with TrueBeam (p<0.001). In free-breathing breast treatment, mean OAR dose decreased from 3.82 to 3.39 Gy (p<0.001), while maximum OAR dose decreased from 25.46 to 24.15 Gy (p=0.039). Smaller but statistically significant differences in mean OAR dose were also observed for head and neck (36.77 versus 37.15 Gy; p=0.036) and prostate with or without nodes (22.07 versus 22.94 Gy; p=0.008).
  • Halcyon significantly reduced treatment delivery time for three treatment categories. Mean time from imaging initiation to completion of the final radiation field decreased from 6.80 to 3.95 minutes for free-breathing breast (p=0.005), from 6.25 to 3.95 minutes for head and neck (p=0.005), and from 4.15 to 3.30 minutes for brain/head (p=0.037). These correspond to approximate relative reductions of 42%, 37%, and 20%, respectively.
  • Other disease-site groups did not demonstrate statistically significant delivery-time differences. Mean treatment time was 3.70 versus 4.75 minutes for chest/abdomen (p=0.084), 4.25 versus 5.20 minutes for prostate with or without nodes (p=0.74), and 6.75 versus 7.60 minutes for palliative treatments (p=0.443), comparing Halcyon with TrueBeam. Although Halcyon was numerically faster across all categories, the observed differences do not establish a consistent efficiency advantage for every treatment site.
  • The investigators developed a treatment allocation decision tree based on dosimetric performance and available machine capabilities. The proposed approach prioritizes free-breathing breast, conventional head and neck, and selected brain/head treatments for Halcyon. Treatments requiring stereotactic delivery, deep-inspiration breath-hold, or six-degree-of-freedom positioning were preferentially assigned to TrueBeam according to the institution's commissioning status. This allocation strategy is specific to the capabilities and clinical workflow of the participating department.
  • The comparison has important planning and methodological limitations. Halcyon plans frequently used three VMAT arcs versus two for TrueBeam, and photon energy was not standardized between systems, creating potential confounding of plan-quality differences. The timing analysis excluded patient entry, positioning preparation, and room turnover, while dose metrics combined heterogeneous organs and prescriptions across anatomical sites. The study was conducted at one institution, did not evaluate clinical toxicity or tumor control, and did not establish actual improvements in departmental throughput or cost-effectiveness.

CLINICAL TAKEAWAY

Halcyon achieved clinically acceptable plan quality and shorter imaging-to-treatment-completion times than TrueBeam for selected treatment sites, particularly free-breathing breast and head and neck radiotherapy. These findings support using local dosimetric and workflow data to guide machine allocation in departments operating multiple linac platforms. However, differences in planning technique, available machine capabilities, and institutional workflow prevent a universal recommendation to prioritize Halcyon over TrueBeam.